![KLS Martin MCO 25 plus Instructions For Use Manual Download Page 43](http://html1.mh-extra.com/html/kls-martin/mco-25-plus/mco-25-plus_instructions-for-use-manual_1979006043.webp)
Revision 5.5
43
5.8.5
Mounting Scanner SOFT FRAC / POWER FRAC
Fractionated scanner operation with MCO 25 plus
is possible if the MCO 25 plus laser is
equipped with the required accessory equipment. Scanner casing and scanner cable are
connected as done for normal scanner operations, see section 5.8.1 “Mounting a Scanner
(Normal Scan)”, page 40.
For fractionated scanning, the scanner handpieces
SOFT FRAC
and
POWER FRAC
are used.
Fig. 5-10:
Fractionated scanner for the MCO 25 plus
A Scanner casing Soft Scan Plus R
B Scanner handpiece
SOFT FRAC
C Scanner handpiece
POWER FRAC
1 Connection thread for mirror joint arm
2 Connection socket for scanner cable to laser surgery device
3 Connection for flushing air hose
4 Beam exit with spacer (tip)
5 Scanner connection thread – scanner handpieces
6 Connection socket for control cable
MICRO POINT R