112
Revision 5.5
3
Check of the required monitoring/safety display and
message system
[ ] 3.1 Shutter (protection against laser beams)
[ ] 3.2 Laser protection goggles
[ ] 3.3 Emission control (acoustic, visual)
[ ] 3.4 Power measuring device (Comparison of internal/external power)
[ ] 3.5 Interlock system (functional check)
[ ] 3.6 ext.
EMERGENCY STOP
switch connection Warning lamps/warnings
(functional check)
4
Electrical safety (IEC 60601-1)
[ ] 4.1 Protective grounding conductor resistance:
≤
0.2
Ω
[ ] 4.2 Ground leakage current:
≤
0.5 mA
5
Measurement of the safety-relevant output parameters
[ ] 5.1 Check calibration with external power meter
(acceptable tolerance: ±20%):
2 W/5 W/10 W/15 W/20 W/ 25 W
WARNING
If there are any safety-relevant problems after the safety check, the device must not be used
any longer.
Measures taken
[ ] Technical Service contacted on: ____________________________________________
[ ] Unit withdrawn from service on: ____________________________________________
[ ] Results entered in medical device logbook on: _________________________________
Person performing
the checks/tests:
Date:
Signature: