6.6.3
Multiple Pulse Mode
When the foot switch is activated, the laser emits a sequence of laser pulses until the foot
switch is released. The display also shows the average laser output (
AVERAGE POWER
) that
is derived from the output power (
MAXIMUM POWER
), the PULSE DURATION and the pause
time (
PAUSE
) or the
PULSE RATE
, respectively.
Parameters in multiple pulse mode
Output power (
MAXIMUM POWER
):
•
The output of the laser is set as in single pulse mode, see section 6.6.2“Single Pulse
Mode”, page 57 or continuous wave mode.
Setting range: 10–25 W in steps of 1 W
Pulse duration:
•
The pulse duration is set as in single pulse mode, see section 6.6.2“Single Pulse Mode”,
page 57.
Setting range: 5 ms–10 s
•
Pulse pause (Pause) Use the
SELECT
key to select the
PAUSE
field.
•
Now increase or decrease the pulse pause value using the arrow keys.
Both keys have an automatic repetition function which increases or decreases the value
gradually when the key is held down. The selected time defines the duration from the end of
one pulse to the beginning of the next.
Setting ranges
From
To
In steps of
5 ms
10 ms
1 ms
10 ms
100 ms
10 ms
0.1 s
1 s
0.1 s
from 1 s
10 s
1 s
Pulse frequency (pulse rate); in
frequency mode
, alternatively the pulse frequency can be
set. This is the number of pulses per second:
Setting ranges
From
To
In steps of
0.1 Hz
1 Hz
0.1 Hz
1 Hz
10 Hz
1 Hz
10 Hz
50 Hz
10 Hz