C-4
APPENDIX C Compass Calibration
HydroChart 3500 Sonar System Operations and Maintenance Manual
P/N 11210087, Rev. 03
6.
Select 115200 from the
Baud Rate
drop-down list box.
7.
Select 40 from the
Output Rate
drop-down list box.
8.
Click
Configure
again to open the Setup dialog box.
9.
Select 115200 from the
Baud Rate
drop-down list box, and then click
OK
.
10.
In the Program dialog box, click
Start
for the X and Y compass components to
start the X an Y level calibration as shown in Figure C-4.
11.
Slowly rotate the table clockwise until the display at the bottom of the dialog box
indicates "......" as shown in Figure C-4, and then press the Space bar to stop the
calibration.
12.
Change the compass orientation so that it is on its side and level to the ground as
shown in Figure C-5.
13.
Click
Start
for the Z compass component to start the Z 90-degree calibration as
14.
Slowly rotate the table clockwise until the at the bottom of the dialog box
indicates "......" as shown in Figure C-6, and then press the Space bar to stop the
calibration.
15.
Disconnect the 5-pin bulkhead connector and remount the compass to the frame.
16.
Reconnect the 5-pin bulkhead connector.
Figure C-4:
Program Dialog Box—X and Y Level Calibration Complete