The TC 1 was checked and controlled in the works according to strictest quality standards. Should however problems arise,
please at first examine the appliance in comparison with the following list.
Disturbance/Operating error Monitoring points/Disposal
9. Fault finding
• Battery was run out, with network adaptor 230 VAC/secondarily 24 VDC
during 14-16 hours charge internal storage batteries of the appliance.
• Sensor resp. measuring cable at sensors of 4 to 20 mA not connected or
defective.
Please check whether the sensor or the cable is being the caus eof the
trouble. Please change both of the parts one after the other.
• Zero point correction was carried out with sensor being under pressure
resp. force. Repeat compensation of zero point correction as described
on page xx or paragragh xx.
• Entry of measuring range was exceeded.
- Short circuit of sensor or cable
- Pressure measuring cell was mechanically overcharged (overstressed).
• Over the external stress socket the batteries of the measuring instrument
can be charged with a Klauke network adaptor (230 VAC, secondarily 24
VDC) or an external voltage of 24 VDC and max. 30 VDC stabilized).
Recommended charging time : 14-16 hours.
• The interface convertzer for the printer has not been connected.
After connection of the device no indication:
Only horizontal lines in the indication of measured values:
Wrong zero point compensation: values too high in the
indication of measured values:
„Over" in the indication:
„Charge battery" in the indication:
„PRN-Adaptor is missing“ in the indication: