-38
-
3.
Measured value storage
3.1
Automatical peak value storage
An automatic peak value storage can be started by pressing the key
for about 2 seconds. All previous peak value storages
are deleted.
The storage of the peak value of a pressing operation is made by a
trigger, provided that the trigger value must have fallen short within the
scope of the pressing cycle so that a new storage can be carried out. By
this, each working cycle which is based on a pressing cycle can be
acquired.
In the notification a bar chart is to be seen on the right side which
signalizes the storage. A maximum of 500 working cycles can be stored.
Theoretically a storage is finished when the bar chart field is filled.
But at any time a storage can be stopped by pressing the key
.
Please observe that only the channel in particular which was selected
before – this might be either force or pressure – is stored. The scanning
rate for the storage of the peak value is 1 ms.
By pressing both of the keys and
the channel indicated in the display is being activated. A scanning rate
of 2 ms is rigidly adjusted and allows a storage of 30.000 measured
values in one minute. With this kind of storage for example a complete
pressing cycle can be stored which later on can be represented
graphically over a printer as run of the curve.
If the peak values of all press-in cycles are to be indicated in the display
you have to press the key .
In case the peak force of each single cycle shall be listed resp.
indicated, this is achieved by a further actuating of the key Output; the
opposite notification appears.
Clear
61.8
kN
65.7
kN
TF 70
F
F ^
61.8
kN
65.7
kN
TF 70
F
F ^
Clear
3.2
Storing without trigger
Prog
ENT
Clear
3.3
Picturing of all single,
stored peak values
Output
F ^ Average value :
63.7
Number of cycles :
68
Minimum value :
61.8
Maximum value :
66.9
61.8
kN
65.7
kN
TF 70
F
F ^
Start of a Recording
End of a recording