Display
Buzzer
Description
Measure
Unit key
required
3
Temperature alarm (high):
n
The upper temperature
warning limit is reached or
exceeded (for example due to
very warm chilled goods or
because the door was open
too long).
n
View and check the tempera-
ture warning limit
.
n
If necessary, correct values of
the temperature warning
limits.
n
Monitor the temperature pro-
gression.
n
If it does not normalise, con-
tact the service department.
n
Transfer chilled goods to new
location.
3
3
Temperature alarm (low)
n
The lower temperature
warning limit is reached or
exceeded (for example after
the door has been opened for
a longer period while the
cooling machine cools down
the unit).
n
Safety equipment triggers,
monitoring circuit has
switched off the cooling
machine.
n
View and check the tempera-
ture warning limit
.
n
If necessary, correct values of
the temperature warning
limits.
n
Monitor the temperature pro-
gression.
n
If it does not normalise, con-
tact the service department.
n
Transfer chilled goods to new
location.
3
Tab. 12: Error messages and status displays of the PC-KIT-STICK
Display
Buzzer
Description
Measure
Unit key
–
Status display:
n
Copying process is running.
n
XX represents the progress of
the copying process in per-
cent.
n
Do not remove USB flash
drive.
–
–
Status display:
n
Copying process is complete.
n
Remove USB flash drive.
–
–
Status display:
n
Memory of the USB flash drive
is full.
n
Use USB flash drive with suffi-
cient memory capacity.
–
–
Error message:
n
The USB flash drive cannot be
read.
n
Use a USB flash drive for-
matted in FAT-32 format.
–
Status displays and error messages
Error messages
27.05.2021
Refrigerator/freezer MED series
62