9
Maintenance
Interval
Maintenance work
At least every 6 months
Check the condenser
Chapter 9.1 “Safety inspection” on page 52
9.1 Safety inspection
Perform safety inspection every two years!
The unit should be inspected by the owner in line
with DGUV regulation 3 (formerly BGV A3).
Subject the unit to a safety inspection at regular
intervals, but at least every two years.
The safety inspection performed by the owner contains the fol-
lowing individual inspections and their documentation:
1 - Visual inspection
2 - Function check
3 - Temperature check
4 - Temperature alarm test
5 - Inspection of the condenser
NOTICE!
If there is any doubt that the unit is in the correct
condition, immediately stop using the unit. To pre-
vent unintended use, label the unit accordingly.
Contact the service department
1.
Check the entire unit for completeness, correct set-up and
damage.
2.
Check the following parts of the unit individually for damage:
n
Door handle
n
Interior
n
Door seal
3.
If there is damage and functionality is not ensured, decom-
mission the unit and contact the service department
.
Contents of the safety inspection
Visual inspection
Maintenance
Safety inspection
27.05.2021
Refrigerator/freezer MED series
52