6.
Turn unit key to position “1”.
ð
The Display shows the current interior temperature.
The set temperature warning limits are
saved automatically. The temperature
warning limits are retained after a power
failure and after the unit has been
switched off.
6.2.3 Adjusting the humidity
MED 288 PRO-ACTIVE
MED 288 ULTIMATE
MED 340 PRO-ACTIVE
MED 340 ULTIMATE
MED 468 PRO-ACTIVE
MED 468 ULTIMATE
MED 520 PRO-ACTIVE
MED 520 ULTIMATE
MED 520 CR PRO-ACTIVE
MED 520 CR ULTIMATE
MED 600 PRO-ACTIVE
MED 720 PRO-ACTIVE
MED 720 ULTIMATE
Two humidity levels are available for the interior of the unit:
high humidity - approx. 90%
low humidity - approx. 65%
If the “high humidity” setting is selected, the recir-
culating blower in the interior is constantly in ope-
ration and is switched off only when the door is
opened. If the “low humidity” setting is selected,
the blower switches off briefly each time the
cooling machine stops.
1.
Insert the unit key in the key switch.
2.
Turn unit key to position “P”.
3.
Press button
[Humidity]
and hold it for at least four bseconds.
4.
Select humidity.
ð
After you select “high humidity”, LED display lights up.
After you select “low humidity”, the LED display goes out.
Overview of units with adjustable
humidity
Adjusting the humidity
Commissioning
Programming the unit > Adjusting the humidity
27.05.2021
Refrigerator/freezer MED series
38