Floating Point Data
PDU address * Parameter
Name
R/W SZ
Mode
Description
10000
U_DEVICE_TYPE
R
U
All
Device type of the sensor (see register IO_DEVICE_TYPE)
10001
U_OPERATIONAL_MODE
R
U
All
Operational mode, normal, service, calibration
10002
U_ERROR_CODE
ErrCode
R
U
All
Most recent/ actual error code (see register IO_ERROR_CODE)
10003
U_STATUS_FLAGS
R
U
All
Device Status flags (see register U_STATUS_FLAGS)
10004
U_BATCH_NR
R
U
All
Production Batch number (see register IO_BATCH_NUMBER)
10005
U_SERIAL_NR
R
U
All
Serial number (see register IO_SERIAL_NUMBER)
10006
FL_SENSOR1_DATA
FSensor1 R
F
All
Temperature compensated (net) radiation sensor 1 in W/m2 with decimal point
10008
FL_STDEV_SENSOR1
FStdDev1 R
F
All
Standard deviation Sensor 1 with decimal point
10010
FL_ SENSOR2_DATA
FSensor2 R
F
All
Temperature compensated Long wave down radiation with decimal point
10102
FL_STDEV_SENSOR2
FStdDev2 R
F
All
Not used, always 0
10014
FL_BODY_TEMPERATURE
R
F
All
Body temperature in ° Kelvin with decimal point
10016
FL_EXT_POWER_SENSOR
R
F
All
External power voltage with decimal point
45
.
A.3 Most important registers
In the table below the most important registers from the real-time processed data are combined.
The FL_XXXX registers are six unsigned 2 word integers and six 32 bit floating points and the floating point take two Modbus
registers with the Most Significant Byte First.
The IO_SCALE_FACTOR (or Resolution in the SmartExplorer software) influences the floating point registers found in this section.
The registers below are implemented from IO_DATAMODEL_VERSION >= 102
By fetching six registers 10000 to 10005 and six floating point registers from 10006 to 10016, all important sensor information
can be collected.