Error reports
PDU address Parameter
R/W
(2)
Type Mode Description
26
IO_ERROR_CODE
R
U16
All
Most recent/ actual error code
27
IO_PROTOCOL_ERROR
R
U16
All
Protocol error/communication error
28
IO_ERROR_COUNT_PRIO1
R
U16
All
Error code priority 1
29
IO_ERROR_COUNT_PRIO2
R
U16
All
Error count priority 2
30
IO_RESTART_COUNT
R
U16
All
Number of controlled restarts
31
IO_FALSE_START_COUNT
R
U16
All
Number of uncontrolled restarts
32
IO_SENSOR_ON_TIME
R
U16
All
On time in seconds (MSB word)
33
IO_SENSOR_ON_TIMEL
R
U16
All
On time in seconds (LSB word)
41
IO_BATCH_NUMBER
R
U16
All
Production batch number
42
IO_SERIAL_NUMBER
R
U16
All
Serial number
43
IO_SOFTWARE_VERSION
R
U16
All
Software version
44
IO_HARDWARE_VERSION
R
U16
All
Hardware version
45
IO_NODE_ID
R
U16
All
(MODBUS®/SMA) device address RS-485
(2)
Writing any value to input registers 26-33 will reset the contents of the registers
Legend
PDU address PDU a 1 = Modbus® register number
Parameter
Name
Name of the register
R/W
Read write
R
Read only
R/W Read/write
Type
Type and size
U16 16 bit unsigned integer
S16 16 bit signed integer
S32
32 bit signed integer (MSB first, LSB last)
Mode
Operation mode
N
available in normal mode
S
available in service mode
C
available in calibration mode (not for users)
F
available in factory mode (not for users)
All
available in all modes
Real-time Data A/D Counts
PDU address Parameter
R/W
Type
Mode Description
18
IO_ADC1_COUNTS
R
S32
All
Input voltage sensor 1 in 0.01 µV
19
(R18=MSB,
R19=LSB)
20
IO_ADC2_COUNTS
R
S32
All
Not supported, always 0
21
22
IO_ADC3_COUNTS
R
S32
All
Input voltage body temperature sensor in 0.01 µV
23
(R22=MSB,
R23=LSB
)
24
IO_ADC4_COUNTS
R
S32
All
Input voltage power sensor in 0.01 µV
25
(R24=MSB,
R25=LSB)
44