PCR-LE2 series 

13

„

„

Single-phase three-wire output or three-phase out-

put

 

Check that the POWER switch is turned off.

 

Check that the breaker of the switchboard is off.

 

Unfasten the screws (two locations) of the OUTPUT 

3P4W(1P3W) terminal block cover, and remove the 

cover.

OUTPUT 3P4W(1P3W) 

terminal block

 

Securely connect the load cables to the OUTPUT 

3P4W(1P3W) terminal block.

If the load has a ground (GND) terminal, be sure to con-
nect it to the G terminal of the PCR27000LE2 OUTPUT 
3P4W(1P3W) terminal block. Be sure to use a wire whose 
diameter is greater than or equal to the diameter of the 
wires used to connect the load.

G

N

U

V

W

G

N

U

V

M8

Three-phase output

Single-phase three-wire output

 

Fasten the cover that you removed in step 3 with the 

screws (two locations).

Connecting the Load (Cont.)

When the load is located at a remote location

Remote control can be used to turn the output off, but it cannot be 
used to turn the POWER switch off. If you want to connect the PCR-
LE2 Series to a load that is located at a remote location, install a 
switch between the OUTPUT terminal block and the load to prevent 
electric shock. Then, turn that switch off.

WARNING

Risk of electric shock.
• Before you install the switch between the OUTPUT ter-

minal block and the load, be sure to turn the POWER 

switch off and turn off the breaker of the switchboard.

• The current rating of the switch must be greater than 

or equal to the maximum current of the PCR-LE2 Se-

ries.

• For the switch circuit, use a multi-pole switch that can 

cut off all lines simultaneously (two poles for single-

phase output, three poles for single-phase three-

phase output, and four poles for three-phase output).

• Be sure to turn the switch off before connecting the 

load to the terminal at the load end of the switch.

• Do not touch the switch terminals when the POWER 

switch is on. Before you connect cables to the OUT-

PUT terminal block, be sure to turn the POWER switch 

off, and then turn off the switchboard.

N   L

N   W   V   U

U

V

W

N

Switch

LOAD

L

N

Switch

LOAD

PCR-LE2

Not used during single-phase three-wire output.

Before you connect the cables, be sure to turn the POWER 

switch off and turn off the breaker of the switchboard.
OUTPUT terminal block

Terminals on the load side

Turn the switch off, and 

then connect the load.

Terminals on the OUTPUT terminal block side

Risk of electric shock. Do not touch the terminals.

Terminals on the load side

Turn the switch off, and then connect the load.

Terminals on the OUTPUT terminal block side

Risk of electric shock. Do not touch the 

terminals.

Connecting wires to the PCR12000LE2, PCR18000LE2, or PCR27000LE2

Summary of Contents for PCR12000LE2

Page 1: ...and waveform can be changed over time Power supply environment testing can be automated In addition to the AC output sequences DC output single phase output or single phase three wire output only and...

Page 2: ...used in the explanations in the manual WARNING Indicates a potentially hazardous situation which if ig nored could result in death or serious injury CAUTION Indicates a potentially hazardous situatio...

Page 3: ...istance Using Power Line Abnormality Simulations In AC mode you can simulate power supply line errors by stop ping the PCR LE2 output and decreasing and increasing the voltage to simulate voltage dips...

Page 4: ...t least 20 cm of space between the air inlet outlet and the wall or obstacles Fix PCR6000LE2s and PCR9000LE2s to the floor using L shaped or other similar brackets Base Hold Angles OP03 KRC are availa...

Page 5: ...block tray in place Stopper 5 Connect the wires and cables to the terminal block and connectors as necessary 6 Return the stopper to its previous position 7 Return the terminal block tray to its prev...

Page 6: ...the switchboard attach crimping terminals that match the screws of the switchboard that has been connected by a qualified engi neer If you will not use one of the optional input power cords prepare a...

Page 7: ...ncy 50 Hz or 60 Hz 2 Check that the POWER switch is turned off 3 Unfasten the screws two locations of the AC INPUT terminal block cover and remove the cover 4 Connect the power cords according to the...

Page 8: ...rs are detected the home position the basic screen appears 0 00 A 0 0 V pk rms AC LOAD 99 00 Hz RANGE L RESP M OUTPUT OFF Turning on the POWER switch on the PCR12000LE2 PCR18000LE2 or 27000LE2 The PCR...

Page 9: ...ber 63 Output on off state If the POWER switch is turned off immediately after the settings have been changed the last settings may not be stored CAUTION Risk of malfunction After turning the POWER sw...

Page 10: ...uirements of single core wires that are used to connect to the load Nominal cross sectional area mm2 AWG reference cross sectional area mm2 Allowable current A at Ta 30 C 86 F 0 9 18 0 82 17 1 25 16 1...

Page 11: ...ill not flow through the PCR LE2 Series even if the POWER switch is turned on Single phase three wire output or three phase out put 1 Check that the POWER switch is turned off 2 Check that the breaker...

Page 12: ...0 mm2 AWG0 2 or less flip the terminal cover upside down and attach it Single phase output 1 Check that the POWER switch is turned off 2 Check that the breaker of the switchboard is off 3 Unfasten the...

Page 13: ...shock Before you install the switch between the OUTPUT ter minal block and the load be sure to turn the POWER switch off and turn off the breaker of the switchboard The current rating of the switch mu...

Page 14: ...s does exceed the limit defined in the SJ T11363 2006 standard The PCR LE2 series is subject to export control laws Trademarks All company names and product names used in the PCR LE2 manual are tradem...

Reviews: