CONFigure Commands
The CONFigure commands are slightly more flexible than the MEASure? commands. These
commands set the parameters, range, and resolution for the specified feature, but they do not
start measurement. To start measurement, use the INITiate or READ? command.
CONF:VOLT:DC
Sets the measurement function to DC voltage measurement and sets the range and resolu-
tion. This command does not start measurement.
Command
CONFigure:VOLTage:DC
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:VOLT:DC:RAT
Sets the measurement function to DC:DC ratio measurement and sets the range and resolu-
tion. This command does not start measurement. In ratio measurement, the specified range
is applied to the input signal, but the auto range is determined by the reference signal.
Command
CONFigure:VOLTage:DC:RATio
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:VOLT:AC
Sets the measurement function to AC voltage measurement and sets the range and resolu-
tion. This command does not start measurement. In AC measurement, the default resolution
is 5 1/2 digits. Therefore, only the front panel display is affected by the specified resolution
parameter.
Command
CONFigure:VOLTage:AC
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:CURR:DC
Sets the measurement function to DC current measurement and sets the range and resolu-
tion. This command does not start measurement.
Command
CONFigure:CURRent:DC
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:CURR:AC
Sets the measurement function to AC current measurement and sets the range and resolu-
tion. This command does not start measurement. In AC measurement, the default resolution
is 5 1/2 digits. Therefore, only the front panel display is affected by the specified resolution
parameter.
Command
CONFigure:CURRent:AC
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:RES
Sets the measurement function to 2-wire resistance measurement and sets the range and
resolution. This command does not start measurement.
Summary of Contents for DME1600
Page 4: ...4 DME1600 This page left blank intentionally...
Page 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Page 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Page 43: ...Appendix A Error Messages B Sample Programs...
Page 55: ......