DME1600
23
Measurement Configuration Commands
Messa
ge Ref
e
re
nce
2
<function>:NPLC
Sets the integration time for the selected measurement function in power line cycles (PLC).
The default setting is “1.” This command only applies to DCV, DCI, 2-wire resistance, and 4-
wire resistance measurements.
Use one of the following character strings to indicate the function name (<function>).
Command
[SENSe:]<function>:NPLCycles {0.02|0.1|1|10|MINimum|MAXimum}
[SENSe:]<function>:NPLCycles? [MINimum|MAXimum]
Response
In response to <function>:NPLC?, the DME1600 returns the integration time for the selected
measurement function.
FREQ:APER
Sets the gate time (or aperture time) for frequency measurement to 10 ms (4 1/2 digits), 100
ms (default setting; 5 1/2 digits), or 1 s (6 1/2 digits). The default setting is “0.1.”
Command
[SENSe:]FREQuency:APERture {0.01|0.1|1|MINimum|MAXimum}
[SENSe:]FREQuency:APERture? [MINimum|MAXimum]
Response
In response to FREQ:APER?, the DME1600 returns the gate time (or aperture time) for
frequency measurement.
PER:APER
Sets the gate time (or aperture time) for period measurement to 10 ms (4 1/2 digits), 100 ms
(default setting; 5 1/2 digits), or 1 s (6 1/2 digits). The default setting is “0.1.”
Command
[SENSe:]PERiod:APERture{0.01|0.1|1|MINimum|MAXimum}
[SENSe:]PERiod:APERture? [MINimum|MAXimum]
Response
In response to FER:APER?, the DME1600 returns the gate time (or aperture time) for period
measurement.
DET:BAND
Sets the minimum input signal frequency. The default setting is “20.” A slow, medium, or fast
AC filter is selected according to the specified frequency.
Command
[SENSe:]DETector:BANDwidth {3|20|200|MINimum|MAXimum}
[SENSe:]DETector:BANDwidth? [MINimum|MAXimum]
Response
In response to DET:BAND?, the DME1600 returns the bandwidth.
AVER:TCON
Sets the digital filter to moving average mode (MOVing) or repeating average mode
(REPeat). The default setting is “MOV.”
Command
[SENSe:]AVERage:TCONtrol {MOVing|REPeat}
[SENSe:]AVERage:TCONtrol?
Response
In response to AVER:TCON?, the DME1600 returns the digital filter mode.
VOLTage:DC
CURRent:DC
RESistance (2-wire resistance)
FRESistance (4-wire resistance)
Summary of Contents for DME1600
Page 4: ...4 DME1600 This page left blank intentionally...
Page 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Page 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Page 43: ...Appendix A Error Messages B Sample Programs...
Page 55: ......