CONFigure Commands
Messa
ge Ref
e
re
nce
2
Command
CONFigure:RESistance
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:FRES
Sets the measurement function to 4-wire resistance measurement and sets the range and
resolution. This command does not start measurement.
Command
CONFigure:FRESistance
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:FREQ
Sets the measurement function to frequency measurement and sets the range and resolu-
tion. This command does not start measurement. In frequency measurement, the DME1600
uses a single range for all input and output signals from 3 Hz to 300 kHz. If no signal is
applied, the measured frequency is zero.
Command
CONFigure:FREQuency
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:PER
Sets the measurement function to period measurement and sets the range and resolution.
This command does not start measurement. In period measurement, the DME1600 uses a
single range for all input and output signals from 0.33 s to 3.3
s. If no signal is applied, the
measured period is zero.
Command
CONFigure:PERiod
{<range>|MIN|MAX|DEF},{<resolution>|MIN|MAX|DEF}
CONF:CONT
Sets the measurement function to continuity measurement. This command does not start
measurement. The range is fixed at 1 k
, and the resolution is fixed at 5 1/2 digits.
Command
CONFigure:CONTinuity
CONF:DIOD
Sets the measurement function to diode testing. This command does not start measurement.
The range is fixed at 1 Vdc with a 1 mA current source, and the resolution is fixed at 5 1/2 dig-
its.
Command
CONFigure:DIODe
CONF:TEMP
Sets the measurement function to RTD temperature measurement. This command does not
start measurement. The resolution is fixed at 6 1/2 digits.
Command
CONFigure:TEMPerature
Summary of Contents for DME1600
Page 4: ...4 DME1600 This page left blank intentionally...
Page 5: ...Setup This chapter explains the settings that you need to configure to use the interfaces 1...
Page 9: ...Message Reference This chapter explains the SCPI com mands 2 Message Reference...
Page 43: ...Appendix A Error Messages B Sample Programs...
Page 55: ......