MT1005

True RMS 

Digital Multimeter

Summary of Contents for MT1005

Page 1: ...MT1005 True RMS Digital Multimeter...

Page 2: ...Contents 1 Introduction 4 2 Safety 4 3 Safety Instructions 8 4 Feature 8 5 Measure Function 11 6 Default Display 19 7 Setup Options 24 8 Specifications 30 2...

Page 3: ...ON Safety This symbol adjacent to another symbol terminal or operating device indicates that the operator must refer to an explanation in the Operating Instructions to avoid personal injury or damage...

Page 4: ...ory appliances Equipment of OVERVOLTAGE CATEGORY II is energy consuming equipment to be supplied from the fixed installation OVERVOLTAGE CATEGORY III Note Examples include switches in a fixed installa...

Page 5: ...d disconnect the power when making resistance or diode tests ALWAYS turn off the power and disconnect the test leads before opening the covers to replace the fuse or batteries NEVER operate the meter...

Page 6: ...f the Meter activate features that augment the function selected using the rotary switch navigate menus or control power to Meter circuits Understanding the Push Buttons 5 RANGE Into manual range and...

Page 7: ...reading 4 Displays measurement information about the input signal 5 Indicates the range the Meter is in and the ranging mode auto or manual 6 Indicates the time set in the internal clock 7 Bluetooth...

Page 8: ...0A Current 20VA overload for 30 seconds on 10 minutes off Input for 0A to 500mA Current measurements Return terminal for all measurements Input for Voltage Continuity Resistance Diode test Conductance...

Page 9: ...tage present Make sure the probe tips are touching the metal contacts inside the outlet before assuming that no voltage is present CAUTION Do not measure AC voltages if a motor on the circuit is being...

Page 10: ...the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive V jack 3 Read the Voltage in the display 1 Set the function switch to the yellow VDC p...

Page 11: ...cur that can damage the meter mV Voltage Measurements 1 Set the function switch to the yellow mV position 2 Press the MODE key Switch to mVDC mVAC 3 Insert the black test lead banana plug into the neg...

Page 12: ...d the Frequency in the display 1 Set the function switch to the yellow CAP position 2 Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the po...

Page 13: ...the red test lead banana plug into the positive jack 4 If the resistance is less than approximately 25 the audible signal will sound If the circuit is open the display will indicate OL 1 Set the funct...

Page 14: ...o the yellow CAP position 2 Press the MODE key Switch to CAP 3 Insert the black test lead banana plug into the negative COM jack Insert the red test lead banana plug into the positive V jack 4 Read th...

Page 15: ...the function switch to the yellow mA position and insert the red test lead banana plug into the A mA jack 4 For current measurements up to 10A DC set the function switch to the yellow 10A position an...

Page 16: ...e A mA jack 3 For current measurements up to 500mA AC set the function switch to the yellow mA position and insert the red test lead banana plug into the A mA jack 4 For current measurements up to 10A...

Page 17: ...Press Graph F1 key the meter will switch to Graph measure Press the START key Press the FAST or SLOW key to adjust the sampling rate Press the CANCEL key to exit Graph and return to normal measuremen...

Page 18: ...f the page In addition the recorded maximum average and minimum values appear in the secondary display with their respective elapsed times To stop a MIN MAX recording session press the STOP key The su...

Page 19: ...t AC measure MIN MAX mode Press the PMAX F4 button To activate the relative mode Press the HOLD REL button for more than 1 second To freeze the display for any function press the HOLD button Press the...

Page 20: ...s the DOWN F3 key to save the selected item press the ENTER F1 key Viewing data stored in the Meter s memory is performed through the SAVE menu Press the DOWN F3 key Position the menu selector next to...

Page 21: ...start recording The recording session will continue until the allocated memory is used the batteries expire the rotary switch is moved The session is terminated by pressing the STOP key Viewing data s...

Page 22: ...s memory is performed through the SAVE menu Press the DOWN F3 key Position the menu selector next to the menu item labeled SETUP and press the ENTER F1 key The Meter s setup options can be reset to de...

Page 23: ...s the ENTER key Using the cursor buttons move the menu selector next to the menu item labeled Numeric Date Time format press the EDIT key to edit select 0 0000 0 0000 MM DD YY DD MM YY and 24 HOUR 12...

Page 24: ...ss the EDIT key To set AUTO POWER OFF Use UP and DOWN to adjust the time to one of the preset values 0 is disable the timeout feature Press the OK key to set the selected time Press the CLOSE key to r...

Page 25: ...by gently prying one end loose then sliding the fuse out of its bracket 4 Install only specified replacement fuses 5 Reinstall the battery door assembly and secure it by turning the screw one half tur...

Page 26: ...test current 0 35mA PEAK Captures peaks 1ms Temperature Sensor Requires type K thermocouple Input Impedance 10M VDC 9M VAC AC Response True RMS Li ion Battery Charge 1 Set the function switch to the...

Page 27: ...ominal Low Battery Indication is displayed if battery voltage drops below operating voltage Battery One 7 4V Li ion battery Fuses mA A ranges 0 8A 1000V ceramic fast blow A range 10A 1000V ceramic fas...

Page 28: ...nge Resolution Accuracy DC Voltage 500mV 1 0 01mV 0 1 5 digits 5V 0 0001V 0 001V 0 01V 0 1V 50V 500V 1000V 0 05 5 digits 0 05 5 digits 0 05 5 digits 0 1 5 digits AC DC 1kHz 5kHz 5V 50V 500V 1000V 0 00...

Page 29: ...3 8 0 5 8 1kHz 5kHz 0 8 5 3 5 500 A 5000 A 50mA 500mA 0 01 A 0 1 A 0 001mA 0 01mA 20A 30 sec max with reduced accuracy AC Current All AC current ranges are specified from 5 of the range to 100 of ran...

Page 30: ...ng 0 0001MHz 0 01 5 Function Range Resolution Accuracy 50Hz 500Hz 5kHz 50kHz 500kHz 5MHz Frequency 0 1 to 99 90 0 01 1 2 reading 2 digits Frequency electronic Sensitivity 2V RMS min 20 to 80 duty cycl...