DC Current Measurements
1. Insert the black test lead banana
plug into the negative
COM
jack.
2. For current measurements up to
5000µA DC, set the function switch
to the yellow
µA
position and insert
the red test lead banana plug into
the
µA/mA
jack.
3. For current measurements up to
500mA DC, set the function switch to
the yellow
mA
position and insert
the red test lead banana plug into
the
µA/mA
jack.
4. For current measurements up to 10A
DC, set the function switch to the
yellow
10A
position and insert the
red test lead banana plug into the
10A
jack.
5. Press the
MODE
key to indicate
"DC" on the display.
6. Read the current in the display.
CAUTION:
Do not make
20A
current measurements for longer than 30
seconds. Exceeding 30 seconds may cause damage to the
meter and/or the test leads.