Performance Verification for Analog Analyzer
3
Keysight U8903B Service Guide
45
1
Connect the function generator to the U8903B input channel 1 (3, 5, or 7) and 2 (4, 6,
or 8) via the Unbalanced connection as shown in
. The BNC-to-banana jack
adapter and the BNC male-to-female T splitter is used for this connection. Preset both
the U8903B and function generator.
Figure 3-4
Function generator to U8903B connection for phase accuracy verification
2
Configure the function generator as follows:
– Output Load: High-Z load
– Output polarity: Normal
– Waveform function: Sine wave
– Amplitude unit = Vrms
– Amplitude offset = 0 V
3
Configure the U8903B input channel 1 (3, 5, or 7) and 2 (4, 6, or 8) as follows:
– Connection type: Unbalanced
– Input bandwidth: 90 kHz
– Input coupling: AC
– Voltage detector type: RMS
– Trigger source: Immediate
– Reference channel: Channel 1 (3, 5, or 7)
– Multi-channel function: Phase
– Input Range:1 V
– Sample size: 131072
4
Set the function generator to produce a sine wave with amplitude of 1 Vrms and
frequency of f
IN
as given in the “
Test record for phase accuracy verification
.
5
Enable the function generator output.
6
Record the phase measurement on the U8903B input channel 2 (4, 6, or 8) in the “
record for phase accuracy verification
.
Unbalanced output
channel 1, 3, 5, 7
U8903B audio
analyzer
Function
generator
Unbalanced output
channel 2, 4, 6, 8
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...