Performance Verification Overview
20
Keysight U8903B Service Guide
Recommended Test Equipment
The recommended test equipment for performance verification tests is listed below. If the
exact equipment is not available, substitute the calibration standards of equivalent
requirement(s).
Table 1-1
Recommended equipment for performance verification tests
Equipment
Recommended model
Critical specification
Digital multimeter
Keysight 3458A
– DC voltage range of 0 V to 10 V or higher
– AC voltage range of 0 Vrms to 10 Vrms or higher
– 8.5-digit resolution or higher
– DC voltage accuracy of 0.3% or higher
– AC voltage accuracy of 0.02% or higher
Calibrator
Fluke 5720A
– DC voltage range of 0 V to 150 V or higher
– AC voltage range of 0 Vrms to 150 Vrms or higher
– Frequency range of 20 Hz to 100 kHz or higher
– DC voltage accuracy of 0.2% or higher
– AC voltage accuracy of 0.02% or higher
Frequency counter
Keysight 53132A Option 010
Keysight 53220A Option 010
– Frequency range of 50 Hz to 80 kHz
– Accuracy of 1 ppm or higher
Oscilloscope
Keysight DSO8064A
Keysight DSO8104A
–
50
Ω
terminator
–
–
Shorting cap (BNC)
–
–
Computer
–
GPIB/USB/LAN and APIB connections
BNC cable
Keysight U8903B-101
1.2 m
XLR cable
Keysight U8903B-102
2 m
BNC-to-banana jack adapter
–
–
BNC-T adapter
–
–
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...