Performance Verification for Analog Generator
2
Keysight U8903B Service Guide
27
Figure 2-2
U8903B to digital multimeter connections for sine wave amplitude accuracy and flatness
verification
2
Configure the digital multimeter as follows:
– Measurement function: AC voltage
– Level filter: Enabled
– NPLC: 10
– Number readings: 1 (with auto sample event)
– Trigger arming: Hold (suspend)
– Trigger mode: Auto
– ACV mode: Synchronous sub-sampled
– Range: Refer to “
Test record for sine wave amplitude accuracy verification
3
Configure the U8903B generator output channels to be tested as follows:
– Connection type: Unbalanced
– Output impedance: 600
Ω
– Output grounding configuration: Floating ground
– Waveform: Sine
– Output state: ON
4
Set both the U8903B output channel 1 and channel 2 to be tested to output a
frequency of 1000 Hz.
5
Set both the U8903B output channel 1 and channel 2 to be tested to output an
amplitude of V
DUT
as given in the “
Test record for sine wave amplitude accuracy
.
6
Set both the digital multimeters’ range as given in the “
amplitude accuracy verification
7
Enable the U8903B output channels.
Unbalanced output
channel 1
U8903B audio
analyzer
Digital multimeter 1
Digital multimeter 2
Unbalanced output
channel 2
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...