Performance Verification for Analog Analyzer
3
Keysight U8903B Service Guide
49
– Function 2: VAC
– Voltage detector type: RMS
– Trigger source: Immediate
– Display view: Graph view
– Graph analysis mode: Magnitude
– FFT windowing: Blackman Harris
– Sample size: 65536
– Graph X-axis spacing: Linear
– Graph trace 1 (3, 5, or 7) State: ON
– Graph trace 2 (4, 6, or 8) State: ON
– Graph Y-axis top limit: -50 dBV
– Graph Y-axis bottom limit: -160 dBV
– Graph marker threshold state: ON
– Graph marker threshold level: -130 dBV
5
On the U8903B analyzer, set the input range as given in the “
.
6
On the calibrator, set the amplitude to 1 V and the frequency of f
IN
.
7
On the U8903B graph analysis mode, set the graph X-axis left and right edge as given
in the “
Test record for CMRR verification
8
On the U8903B graph analysis mode, set the graph Marker 1 state to ON and assign it
to Trace 1 (3, 5, or 7). Set the graph Marker 2 state to ON and assign it to Trace 2 (4, 6,
or 8).
9
Enable the calibrator output.
10
Using the graph Marker 1 search function, place Marker 1 at the highest amplitude
acquired from the graph measurement.
11
Record the Marker 1’s X-value measurement as F
CMRR
, multiply Marker 1’s Y-value
measurement with –1 and record as CMRR on the U8903B in the “
.
12
for the Marker 2 measurement.
13
for the rest of the frequency of f
IN
.
14
for the rest of the U8903B analyzer input range.
15
to complete the verification of the other U8903B input
channels.
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...