4
Performance Verification for Digital Generator and Digital Analyzer
64
Keysight U8903B Service Guide
– Digital output sampling rate: 48 kHz
– Digital output function: SINE waveform
– SINE waveform frequency: 1 kHz
– SINE waveform amplitude: 1 FFS
– Digital output state: ON
4
Verification steps for word length at 8 bits:
a
Set the DSI audio resolution to 8 bits.
b
Set the DSI audio word length to 8 bits.
c
Set the DSI Master Clock multiplier to 128.
d
Set the digital output sampling rate to 16 kHz.
e
Set the DSI audio resolution based on the Word Length as given in “
DSI master clock output frequency accuracy verification
f
Set the DSI audio word length to 32 bits.
g
Set the Digital Output Sampling Rate based on the sampling rate (Hz) as given in
“
Test record for DSI master clock output frequency accuracy verification
.
h
Set the DSI Master Clock Multiplier based on the Master Clock Multiplier as given in
“
Test record for DSI master clock output frequency accuracy verification
.
i
Measure the Master Clock Rate twice and average the two measurements.
j
Record the average Master Clock Rate, F
measure
clock output frequency accuracy verification
.
k
Compute the master clock error using the following equation. MCLK is the master
clock rate calculated where MCLK = MUL
×
Sampling Rate (Hz).
Δ
F
error
ppm = [(F
measure
– MCLK) / MCLK]
×
1000000
5
Verification steps for word length at 12 bits:
a
Set the DSI audio resolution to 12 bits.
b
Set the DSI audio word length to 12 bits.
c
Set the DSI Master Clock multiplier to 96.
d
Set the digital output sampling rate to 16 kHz.
e
Set the DSI audio resolution based on the Word Length as given in “
DSI master clock output frequency accuracy verification
f
Set the DSI audio word length to 32 bits.
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...