Performance Verification for Analog Analyzer
3
Keysight U8903B Service Guide
39
Performance Verification
DC accuracy verification
This test verifies the DC accuracy of the U8903B analyzer by measuring a range of DC
voltage amplitudes supplied by a calibrator to ensure that it is within the published
specification.
1
Connect the calibrator to the U8903B input channel to be tested via the Unbalanced
connection as shown in
. The BNC-to-banana jack adapter and the BNC
male-to-female T splitter is used for this connection. Preset both the U8903B and
calibrator.
Figure 3-1
Calibrator to U8903B connection for DC accuracy verification
2
Configure the calibrator as follows:
– Output state: Standby
3
Configure the U8903B input channels, channel 1 (3, 5, or 7) and channel 2 (4, 6, or 8) as
shown in
to be tested as follows:
– Connection type: Unbalanced
– Input bandwidth: 90 kHz
– Input coupling: DC
– Voltage detector type: RMS
Table 3-1
DC accuracy specification
DC accuracy specification
Equipment used
±
1%
Calibrator
NOTE
Power on the U8903B audio analyzer for at least 30 minutes before performing the
performance verification.
Unbalanced input
channel 1, 3, 5, 7
U8903B audio
analyzer
Calibrator
Unbalanced input
channel 2, 4, 6, 8
Summary of Contents for U8903B
Page 1: ...Keysight U8903B Performance Audio Analyzer Service Guide ...
Page 12: ...12 Keysight U8903B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 126: ...A Appendix A 126 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 160: ...B Appendix B 160 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 164: ...C Appendix C 164 Keysight U8903B Service Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...