4
System-Related Operation
Remote Operation
150
U3606B User’s Guide
Remote Operation
The U3606B is shipped with both a GPIB (IEEE- 488)
interface and a USB 2.0 interface on the rear panel. Only
one interface can be enabled at a time. The GPIB interface
is selected by default when the U3606B is shipped from the
factory.
The remote interface can be selected from the front- panel
only.
•
The interface selection is stored in the non- volatile
memory, and does not change when power is turned off
or after a remote interface reset.
•
If you select the GPIB interface, you must select a unique
address for the U3606B. The current address for the
U3606B is displayed on the lower secondary display in
the Utility menu.
•
You can select between two USB class: USB- TMC or
USB- CDC. USB- TMC is the default USB 2.0 full- speed
communication protocol interface that is compliant to
USB standards, while USB- CDC is simulates a serial
communication (RS- 232) interface at the PC through the
physical USB port connection.
The instrument automatically enters the remote state
whenever SCPI commands are received over the GPIB or
USB interface. The Rmt annunciator is illuminated and the
front panel keys are locked when in the remote state. Press
[Local]
to return the U3606B to front panel operation.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...