System-Related Operation
4
Utility Menu Items
U3606B User’s Guide
135
To set the number of points, press
[
]
until the cursor
position is on the number of points on the primary display.
Press
[
]
or
[
]
to change the cursor position and
[
]
or
[
]
to change the values for the number of points (2 to
1999).
N O T E
•
The range of values for the number of points is from 2 to 1999, with the
default value at 10.
•
The number of readings used to perform the smooth function
calculation depends on the number of points set. For example, if the
number of points is set to 10 (N = 10), the smooth function will only use
a maximum of 10 readings for averaging instead of an incremental
value as used in the average function of the MinMax operation. When
the instrument reaches the 11th reading or more, it will take the
previous averaged value multiply by (N–1), and the current value to
perform the averaging. Note the difference given below between the
smooth function and average function for a x number of readings:
Smooth function
Average function
V
1
= r
1
V
1
= r
1
V
2
= (r
1
+ r
2
)/2
V
2
= (r
1
+ r
2
)/2
V
3
= (r
1
+ r
2
+ r
3
)/3
V
3
= (r
1
+ r
2
+ r
3
)/3
…
…
V
10
= (r
1
+ r
2
+ … + r
10
)/10
V
10
= (r
1
+ r
2
+ … + r
10
)/10
V
11
= (V
10
*9 + r
11
)/10
V
11
= (r
1
+ r
2
+ … + r
10
+ r
11
)/11
V
12
= [(V
11
* 9) + r
12
]/10
V
12
= (r
1
+ r
2
+ … + r
11
+ r
12
)/12
…
…
V
x–1
= [(V
x–2
* 9) + r
x–1
]/10
V
x–1
= [r
1
+ r
2
+ … + r
x–2
+ r
x–1
]/(x–1)
V
x
= [(V
x–1
* 9) + r
x
]/10
V
x
= [r
1
+ r
2
+ … + r
x–1
+ r
x
]/x
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...