5
Characteristics and Specifications
Digital Multimeter Specifications
172
U3606B User’s Guide
Low-resistance
Measurement method:
Four-wire, the test current is sent from the FORCE
terminals and the resistance is measured by the
SENSE terminals.
Input protection:
•
FORCE terminals: Protected with a 3.15 A/250 V
FF fuse
•
SENSE terminals: 1000 V
rms
on all ranges, < 0.3 A
short circuit
Open circuit voltage:
< +8.6 VDC
Continuity
Measurement method:
0.83 mA ± 0.2% constant current source
Open circuit voltage:
< +5.0 V
dc
Audible tone:
Continuous beep when reading is less than
threshold resistance of 10
Ω
at 1.0 k
Ω
range
Input protection:
1000 V
rms
on all ranges, < 0.3 A short circuit
Diode
Measurement method:
0.83 mA ± 0.2% constant current source
Open circuit voltage:
< +5.0 V
dc
Audible tone:
•
Continuous beep when level is below +50 mV DC
•
Single tone for normal forward-biased diode or
semiconductor junction where 0.3 V
¤
reading
¤
0.8 V
Input protection:
1000 V
rms
on all ranges, < 0.3 A short circuit
Capacitance
Measurement method:
Computed from constant current source charge
time, typical 0.2 V to 1.4 V signal level
Maximum voltage at
full scale:
•
For 1 nF to 10
μ
F range: < 1.5 V
•
For 100
μ
F to 10000
μ
F: < 0.33 V
Input protection:
1000 V
rms
on all ranges, < 0.3 A short circuit
Response time:
Approximately 1 s for 100
μ
F and ranges below 100
μ
F
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...