System-Related Operation
Utility Menu Items
142
U3606B User’s Guide
Selecting a dBm reference resistance value
The dBm function is logarithmic, and is based on a
calculation of power delivered to a reference resistance,
relative to 1 mW. The following procedure shows you how to
select an appropriate dBm reference resistance value. The
factory default dBm reference resistance value is 600
Ω
.
Press
[
]
or
[
]
to change the dBm reference resistance
value (0001
Ω
to 9999
Ω
).
Performing a self-test
A power- on self- test occurs immediately when you turn on
the instrument. This limited test assures you that the
U3606B is operational.
The following procedure shows you how to perform a more
extensive self- test. A complete self- test performs a series of
internal tests, and may take up to 30 seconds to complete.
For remote interface operation, refer to the
*TST
command
in the
U3606B Programmer’s Reference
.
C A U T I O N
Before performing the complete self-test, ensure that all test leads are
removed from the input and output terminals.
Summary of Contents for U3606B
Page 1: ...Keysight U3606B Multimeter DC Power Supply User s Guide ...
Page 2: ...THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 3: ...U3606B User s Guide I ...
Page 16: ...XIV U3606B User s Guide THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 25: ...Introduction 1 Preparing the U3606B U3606B User s Guide 7 Adjust the carry handle 1 2a 3 2b ...
Page 27: ...Introduction 1 Preparing the U3606B U3606B User s Guide 9 Figure 1 3 Rack mount dimensions ...
Page 120: ...3 DC Power Supply Operation Square Wave Operation 102 U3606B User s Guide Enable the output 7 ...