Operation and Features
Triggering the Multimeter
78
U3606B User’s Guide
Remote interface operation:
•
The
TRIGger:SOURce BUS
command selects the bus
trigger source.
•
The
MEASure?
command overwrites the bus trigger and
triggers the U3606B and returns a measurement.
•
The
READ?
command does not overwrite the bus trigger,
and if selected, generates an error. It will only trigger the
instrument and return a measurement when the
IMMEdiate
trigger is selected.
•
The
INITiate
command only initiates the measurement
and needs a trigger (
*TRG
command) to make the actual
measurement.
Refer to the
U3606B Programmer’s Reference
for the syntax
and complete descriptions for these commands.