N7026A Probe
N7026A 150 MHz AC/DC High Sensitivity Current Probes User’s Guide
21
4
Perform the steps described in
“Performing Demagnetization and Zero Offset"
on page 26.
Always carry out demagnetization before making measurements.
5
Pull the probe’s opening lever to open the sensor head. The UNLOCK indication is displayed
on the sensor head. Keep sliding the lever backwards into the Open position.
6
With the probe’s sensor head around the insulated conductor to be measured, slide the
opening lever of the probe
towards the front of the sensor head
into the
Lock
position. While
doing so, ensure that the
UNLOCK
indication disappears which indicates that the sensor
head is properly closed. If the sensor head is not closed properly, accurate measurements
will not be possible.
Figure 4
Use the Opening Lever
External power supply
plugged into N7026A
CAUTION
Always use the opening lever when opening/closing the probe’s
sensor head. If the upper core is forced to open when the sensor head
is locked, the open-close mechanism can be damaged.
Slide the lever forward until
the UNLOCK indication disappears
Do Not Press Here!
Upper Core