30
N7026A 150 MHz AC/DC High Sensitivity Current Probes User’s Guide
N7026A Probe
Performance Verification for N7026A
Use the following procedure to test the warranted Accuracy specifications for the N7026A
probe (as listed on
page 10
). The recommended test interval to warrant performance of this
probe is once a year. However, you can also test the probe’s performance as and when
required using the recommended test equipment and by following the procedure
documented in this chapter.
Procedure
1
Turn on the oscilloscope and then connect the N7026A probe to the oscilloscope.
2
Set A/div to
10mA/div
.
3
On an Infiniium oscilloscope, choose
Setup > Probe Configuration....
to access the
Probe
Configuration
dialog box and then record the
Attenuation
scale factor. On an InfiniiVision
oscilloscope, use th
e
Probe
softkey to record the attenuation scale factor of the probe.
Table 6
Required Test Equipment
Description
Minimum Requirements
Recommended Test Equipment
Oscilloscope
Bandwidth
≥
250 MHz
Amplitude accuracy:
≤
0.4%
Infiniium or InfiniiVision with
1 M
Ω
input
Two Digital Multimeters
-
Keysight U1251B Handheld
Digital Multimeter or equivalent
DVM/DMM
AC Current Generator
AC 10 A
rms
, 50 or 60 Hz, sine
wave
Amplitude accuracy:
≤
0.3%
Keysight AC6800 or Wavetek
9100
Probe Adapter
-
Keysight N1022B Probe Adapter
Power Supply
-
Keysight 1143A Probe Offset
Control and Power Module
Test Cables (with banana jacks) -
-
BNC Cable
-
-
BNC (F) to Banana Adapter
-
-
NMD 3.5mm (M) to 3.5mm (F)
adapter
-
-
3.5mm (M) to BNC (F) adapter
-
-