N7026A Probe
N7026A 150 MHz AC/DC High Sensitivity Current Probes User’s Guide
31
The
attenuation scale factor
should be
1 ±10%
. If the scale factor is beyond 1
±
10%, then return the probe for service. Examples of correct attenuation scale
factor are shown below.
Figure 8
Attenuation scale factor on an Infiniium oscilloscope
Figure 9
Attenuation scale factor on an InfiniiVision oscilloscope
4
Remove the N7026A probe from the oscilloscope.
5
To provide power to the N7026A probe, connect the power cord of the N1022B probe
adapter to the Power connector on the front panel of the 1143A power module. Then
connect the N7026A probe’s output to the N1022B probe adapter’s input.