Keysight N5990A-155 DisplayPort User Guide
39
Connection Diagrams
4
Calibration Procedures
Rise Fall Time Measurement Calibration
Purpose and Method
This calibration measures the Rise Fall Time.
For this purpose, the Scope Rise time measurement function on the
real-time scope is used. All jitter components are disabled and the pattern
is set to the D24_3. Then, the 20%-80% transition time is measured and
averaged over a number of cycles.
This calibration is only available for HBR3 data rate.
Connection Diagram
Figure 13
Connection Diagram for TP1 Calibration
Parameters in Expert Mode
• Calibration
•
Oscilloscope Bandwidth
•
Measurement Cycles
• Pattern
•
Sequence File
• Voltage
•
Differential Voltage
The description for these parameters are listed in