Electronic Calibration Modules Reference Guide 3-15
Operating ECal Module
ECal Troubleshooting
Table 3-5
N755xA Troubleshooting
LED Color
Description
Possible Causes
Solutions
Orange
ECal is connected
and functioning, but
is not
communicating with
driver software.
ECal module is connected to an
instrument (or PC) without the
appropriate driver software installed.
Connect the ECal module to a
compatible instrument.
Ensure that the instrument software is
up to date.
Faulty instrument USB port
Connect to a different USB port on the
instrument.
Faulty USB cable
Connect via a different USB cable.
Faulty USB hub
Connect ECal directly to a USB port on
the instrument.
Off
The ECal is not
functioning.
ECal is not connected to instrument.
Double check the USB connectors are
firmly inserted.
Faulty instrument USB port
Connect to a different USB port on the
instrument.
Faulty USB cable
Connect via a different USB cable.
Faulty USB hub
Connect ECal directly to a USB port on
the instrument.
Faulty ECal module
Return to Keysight for service. Refer to
.