![Keysight N4431D Reference Manual Download Page 100](http://html1.mh-extra.com/html/keysight/n4431d/n4431d_reference-manual_1961141100.webp)
5-10 Electronic Calibration Modules Reference Guide
Specifications and Characteristics
Characteristic Performance
5-
Characteristic Performance
Characteristic performance for RF and microwave ECal modules are provided
in the following tables, which describe non-warranted performance that most
units exhibit.
N755xA Series
Table 5-17 Characteristic Performance for N755xA (3.5 mm)
a
a. When applied power exceeds –10 dBm, calibration results will be degraded from the performance indicated in this table.
Parameter
Frequency Range
DC to
500 MHz
500 MHz
to 4 GHz
4 to 6.5
GHz
6.5 to 9
GHz
9 to 14
GHz
14 to 18
GHz
18 to 26.5
GHz
Transmission Tracking (±dB)
b
b. Values based on using the network analyzer N5234A Option 200.
0.15
0.16
0.22
0.22
0.30
0.30
0.35
Load Match (dB)
34
29
28
22
26
26
24
Table 5-18 Characteristic Performance for N755xA (Type-N 50
Ω
)
a
a. When applied power exceeds –10 dBm, calibration results will be degraded from the performance indicated in this table.
Parameter
Frequency Range
DC to 500
MHz
500 MHz
to 4 GHz
4 to 6.5
GHz
6.5 to 9
GHz
9 to 14 GHz 14 to 18
GHz
Transmission Tracking (±dB)
b
b. Values based on using the network analyzer N5234A Option 200.
0.15
0.16
0.22
0.22
0.30
0.30
Load Match (dB)
34
29
28
22
26
26