Step 6 - Verifying Operation of the M8920A
Auto Align
Before verifying operation, you will need to perform an auto alignment. This should be
done,
at least once a week
any time a module is switched out of the chassis
any time a connection change has been made
whenever the ambient temperature has change ±5°C
1. Start the Modular TRX application by selecting the desktop icon.
2. Select the hardware configuration for the Radio Test Solutionthen select Run
Selected.
(Located in the upper right corner of the display.)
3. At the top of the display, select the Screen tab to open the
Mode/Measurement/View Selector dialog.