72 Language Dictionary
Description Of Subsystem Commands
Figure 7-2 is a tree diagram of the subsystem commands. Commands followed by a question mark (?) take only the query
form. Except as noted in the syntax descriptions, all other commands take both the command and query form. The
commands are listed in alphabetical order and the commands within each subsystem are grouped alphabetically under the
subsystem.
Figure 7-2. Subsystem Commands Tree Diagram
ABOR
This command cancels any trigger actions presently in process. Pending trigger levels are reset equal to their corresponding
immediate values.
ABOR
also resets the WTG bit in the Operation Condition Status register (see "Chapter 8 - Status
Reporting"). If
INIT:CONT ON
has been programmed, the trigger subsystem initiates itself immediately after
ABORt,
thereby setting WTG.
ABOR
is executed at power turn on and upon execution of
*RCL
or
RST.
Command Syntax
ABORt
Parameters
(None)
Examples
ABOR
Query Syntax
(None)
Related Commands
INIT *RST *TRG TRIG
Calibration Commands
See Appendix B
Summary of Contents for E4356A
Page 1: ...Operating and Programming Guide Keysight Model E4356A Telecommunications DC Power Supply ...