84 Language Dictionary
Command Syntax
[SOURce]:VOLTage[:LEVel][:IMMediate][AMPLitude] <NRf+>
[SOURce][:VOLTage[:LEVel]:TRIGgered[:AMPLitude] <NRf+>
Parameters
Table 7-1
Default Suffix
V
*RST Value
Table 7-1
Examples
VOLT 200
MA
VOLTAGE:LEVEL 200 MA
VOLTAGE:LEVEL:IMMEDIATE:AMPLITUDE 2.5
VOLT:TRIG 20
VOLTAGE:LEVEL:TRIGGERED 20
Query Syntax
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]?
MAX
[SOURce]:VOLTage[:LEVel][:IMMediate][:AMPLitude]? MIN
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]?
[SOURce]:VOLTage[LEVel]:TRIGgered[:AMPLitude]? MAX
[SOURce]:VOLTage[:LEVel]:TRIGgered[:AMPLitude]? MIN
Returned Parameters
<NR3>
VOLT?
and
VOLT:TRIG?
return presently programmed
immediate and triggered levels. If not triggered level is programmed,
both returned values are the same.
VOLT? MAX
and
VOLT? MIN
return the maximum and
minimum programmable immediate voltage levels.
VOLT:TRIG? MAX
and
VOLT:TRIG? MIN
return the
maximum and minimum programmable triggered voltage levels.
Related Commands
For
VOLT
*SAV *RCL
*RST
For
VOLT:TRIG
ABOR VOLT *RST
VOLT:PROT
This command sets the overvoltage protection (OVP) level of the power supply. If the output voltage exceeds the OVP
level, then the power supply output is disabled and the Questionable Condition status register OV bit is set (see "Chapter 8 -
Status Reporting"). An overvoltage condition can be cleared with the
OUTP:PROT:CLE
command after the condition
that caused the OVP trip is removed. The OVP always trips with zero delay and is unaffected by the
OUTP:PROT:DEL
command.
Command Syntax
[SOURce]:VOLTage:PROTection[:LEVel] <NRf+>
*Alternate
Syntax
[SOURce]:VOLTage:PROTection:AMPLitude <NRf+>
Parameters
Table 7-1
Default Suffix
V
*RST Value
MAX
Examples
VOLT:PROT 21.5 VOLT:PROT:LEV MAX
VOLTAGE:PROTECTION:LEVEL 145E-1
Query Syntax
[SOURce]:VOLTage:PROTection[:LEVel]?
[SOURce]:VOLTage:PROTection [:LEVel]? MIN
[SOURce]:VOLTage:PROTection [:LEVel]? MAX
Returned Parameters
<NR3>
VOLT:PROT?
returns presently programmed
OVP level.
VOLT:PROT? MAX
and
VOLT:PROT? MIN
return the
maximum and minimum programmable OVP levels.
Related Commands
OUTP:PROT:CLE *RST *SAV *RCL
* Available to accommodate earlier power supply programs.