5
Service Information
52
Keysight E3630A Operating and Service Manual
17
Reduce the value of R
L
gradually while observing the output current indicated
by the DVM. The current should increase to a maximum of 2.75 A ±5%
(2.6125 A to 2.8875 A) before it begins to decrease.
18
Connect a short across R
L
and then recheck the current indicated by the DVM.
The short-circuit current of this output should be 1 A ±15% (0.85 A to 1.15 A).
Disconnect the test setup from the supply.
Figure 5-4
Output current, test set up
Load regulation (load effect)
Definition:
The immediate change, •E
OUT
, in the static value of DC output
voltage resulting from a change in load resistance from open circuit to the value
that yields maximum rated output current (or vice versa).
To check the load regulation:
1
Connect the test equipment across the output of the +20 V supply as shown in
. Operate the electronic load in constant current mode and set its
current to the full rated value of the +20 V supply. Input off the electronic load.
Summary of Contents for E3630A
Page 1: ...Triple Output Power Supply Keysight Model E3630A Operating and Service Manual ...
Page 10: ...10 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 64: ...6 Component Locator Diagrams 64 Keysight E3630A Operating and Service Manual Main Board Top ...
Page 68: ...6 Component Locator Diagrams 68 Keysight E3630A Operating and Service Manual 7 Segment Top ...
Page 69: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 69 7 Segment Bottom ...