Service Information
5
Keysight E3630A Operating and Service Manual
59
Troubleshooting
Before attempting to troubleshoot the power supply, ensure that the fault is with
the supply and not with an associated piece of equipment. You can determine this
without removing the covers from the power supply by using the appropriate
portions of the
paragraph.
Open fuse troubleshooting
Although transients or fatigue can cause a fuse to blow, it is a good idea to
inspect the unit for obvious shorts such as damaged wiring, charred components,
or extraneous metal parts or wire clippings in contact with circuit board
conductors before replacing the fuse. The rating of the correct replacement fuse
depends on the line voltage option of the instrument: for Option OE3, use a
slow-blow 1.0-amp fuse and standard and Option OE9, use a slow-blow 1.6-amp
fuse.
CAUTION
Before applying power to the supply, make certain that its line voltage
selector switch (S2) is set for the line voltage to be used.
Summary of Contents for E3630A
Page 1: ...Triple Output Power Supply Keysight Model E3630A Operating and Service Manual ...
Page 10: ...10 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 64: ...6 Component Locator Diagrams 64 Keysight E3630A Operating and Service Manual Main Board Top ...
Page 68: ...6 Component Locator Diagrams 68 Keysight E3630A Operating and Service Manual 7 Segment Top ...
Page 69: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 69 7 Segment Bottom ...