Service Information
5
Keysight E3630A Operating and Service Manual
61
3
Turn on the supply and adjust R5 on the display board until front panel AMPS
display reads exactly DVM value divided by R
S
.
4
To calibrate the ammeter for +6 V output, 6 V meter switch and turn
+6 V VOLTAGE control fully clockwise.
5
Adjust R61 on the main board until front panel AMPS display reads exactly
DVM value divided by R
S
.
Summary of Contents for E3630A
Page 1: ...Triple Output Power Supply Keysight Model E3630A Operating and Service Manual ...
Page 10: ...10 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 64: ...6 Component Locator Diagrams 64 Keysight E3630A Operating and Service Manual Main Board Top ...
Page 68: ...6 Component Locator Diagrams 68 Keysight E3630A Operating and Service Manual 7 Segment Top ...
Page 69: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 69 7 Segment Bottom ...