Stability (Drift)
Definition:
The change in output voltage (DC to 20 Hz) for the first 8 hours
following a 30-minute warm-up period with constant input line voltage, constant
load resistance and constant ambient temperature.
To measure the stability:
1
Connect the test equipment across the output of the +20 V supply as shown in
.
2
Operate the electronic load in constant current mode and set its current to the
full rated value of power supply.
3
Turn on the supply.
4
Turn up output voltage to the full rated value as read on the digital voltmeter.
5
After a 30-minute warm-up, note the voltage on DVM.
6
The output voltage reading should deviate less than 0.1% plus 5 mV from the
reading obtained in step e over a period of 8 hours.
7
Repeat for the remaining supply outputs.
Summary of Contents for E3630A
Page 1: ...Triple Output Power Supply Keysight Model E3630A Operating and Service Manual ...
Page 10: ...10 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 14: ...14 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 16: ...16 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 18: ...18 Keysight E3630A Operating and Service Manual THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK ...
Page 64: ...6 Component Locator Diagrams 64 Keysight E3630A Operating and Service Manual Main Board Top ...
Page 68: ...6 Component Locator Diagrams 68 Keysight E3630A Operating and Service Manual 7 Segment Top ...
Page 69: ...Component Locator Diagrams 6 Keysight E3630A Operating and Service Manual 69 7 Segment Bottom ...