68
Getting Started Guide
3
Test Set Operating System
Using the test set recovery system
Step
Notes
1
Make sure the test set is turned off.
2
Turn on the test set.
•
Press the down arrow key to move the
highlight to
Agilent Recovery System
,
then press
Enter
.
After the Agilent Technologies screen is
displayed,
this screen is displayed for five seconds.
3
When the Agilent Recovery System has
booted, follow the on-screen instructions
to recover the image of the C: drive.
•
Press
2
, then press
Enter
to select the
recovery.
•
Press
1
, then press
Enter
to continue.
•
Press
1
, then
Enter
to confirm.
It may take up to 25 minutes for this process to
complete
After exiting the Agilent Recovery System, the test set reboots. If the original factory test set
system has been restored, the test set re-executes the following process:
Additional recovery steps may be required to fully recover the system
to a more current working state. This could involve restoring your own
backups of the test set configuration, including re-installing
applications, data, and performing system customizations.