Verification 19
Averaging the CC Measurements
The CC Load Effect and CC Source Effect tests measure the dc regulation of the power supply's output current. When
doing these tests, you must be sure that the readings taken are truly dc regulation values and not instantaneous ac peaks of
the output current ripple. You can do this by making each measurement several times and then using the average of the
measurements as your test value. Voltmeters such as the Keysight 3458A System Voltmeter can be programmed to take
just such statistical average readings as required by these tests.
The following steps show how to set up the voltmeter from its front panel to take a statistical average of l00 readings.
represents the unlabeled shift key in the FUNCTION/RANGE group.
1.
Program 10 power line cycles per sample by pressing
.
2.
Program 100 samples per trigger by pressing
.
3.
Set up voltmeter to take measurements in the statistical mode as follows:
a.
Press
.
b.
Press
until MATH function is selected, then press
.
c.
Press
until STAT function is selected, then press
.
4.
Now set up voltmeter to read the average of the measurements as follows:
a.
Press
.
b.
Press
until RMATH function is selected, then press
.
c.
Press
until MEAN function is selected, then press
.
5.
Execute the average reading program by pressing
.
6.
Wait for 100 readings and then read the average measurement by pressing
.
Record this as your result.
Summary of Contents for 669 A Series
Page 2: ...Service Manual Keysight Series 669xA GPIB DC Power Supplies ...
Page 3: ......
Page 27: ......
Page 56: ...Troubleshooting 53 Figure 3 15 3 Inch Front Panel Frame Assembly ...
Page 57: ...54 Troubleshooting Figure 3 16 Assembly A10 Exploded View ...
Page 58: ...Troubleshooting 55 Figure 3 17 Assembly A10 Exploded View 6690A ...
Page 59: ...56 Troubleshooting Figure 3 18 Assembly A10 Exploded View 6691A 6692A ...
Page 60: ...Troubleshooting 57 Figure 3 19 Three Phase Line Choke Subchassis Wiring ...
Page 61: ...58 Troubleshooting Figure 3 20 24 Volt Fan Transformer ...
Page 77: ...74 Diagrams Figure 6 1 Test Point Waveforms for Table 6 3 sheet 2 of 2 ...
Page 79: ...Figure 6 3 A1 Front Panel Board Assembly Diagram ...
Page 81: ...Figure 6 5 A2 GPIB Board Component Location ...
Page 83: ...Figure 6 7 A3 FET Board Component and Test Point Location ...
Page 87: ...Figure 6 11 A4 AC Input Board Component and Test Point Location 12 9 10 11 ...
Page 89: ...Figure 6 13 A5 DC Rail Board Component and Test Point Location 13 14 ...
Page 91: ...Figure 6 15 A6 Bias Board Component and Test Point Location 15 16 17 18 19 19 20 22 ...
Page 94: ...Figure 6 17 Power Mesh Schematic Diagram All Models ...
Page 103: ......