Troubleshooting 25
3
Troubleshooting
Introduction
Shock Hazard
: Most of the procedures in this chapter must be performed with power applied and
protective covers removed. These procedures should be done only by trained service personnel aware of
the hazard from electrical shock.
This instrument uses components that can be damaged or suffer serious performance degradation due to
ESD (electrostatic discharge). Observe standard antistatic precautions to avoid damage to the
components (see Chapter 1).
This chapter provides troubleshooting and repair information for the power supply. Before beginning troubleshooting
procedures, make certain the problem is in the power supply and not with an associated circuit, the GPIB controller (for
GPIB system power supplies), or ac input line. Without removing the covers, you can use the Verification tests in Chapter 2
to determine if the power supply is operating normally.
The information in this chapter is organized as follows:
Topic
Information Given
Test Equipment Required
Equipment required for completing all the tests in this chapter.
Troubleshooting Procedures
An explanation of the error codes and messages generated during the power-on selftest.
A series of flow charts for systematic location of defective boards, circuits, and
components.
Post-Repair Adjustments
Calibration and EEPROM initialization procedures required after the replacement of
certain critical components.
Disassembly Procedures
Gaining access to and/or replacing components.
Test Equipment Required
Table 3-1. Test Equipment Required
Equipment
Purpose
Recommended Model
Test Clips
To gain access to IC pins.
AP Products No. LTC
Ammeter/Current Shunt
To measure output current.
Keysight 6690A: Burster 1280
Keysight 6691A, 6692A: Guildline 9230/300
Oscilloscope
To check waveforms and signal levels. Keysight 54504A
GPIB Controller
To communicate with power supply
via the GPIB (for system units).
Keysight BASIC series
DC Voltmeter
To measure output voltage and
current, bias and references.
Keysight 3458A
Summary of Contents for 669 A Series
Page 2: ...Service Manual Keysight Series 669xA GPIB DC Power Supplies ...
Page 3: ......
Page 27: ......
Page 56: ...Troubleshooting 53 Figure 3 15 3 Inch Front Panel Frame Assembly ...
Page 57: ...54 Troubleshooting Figure 3 16 Assembly A10 Exploded View ...
Page 58: ...Troubleshooting 55 Figure 3 17 Assembly A10 Exploded View 6690A ...
Page 59: ...56 Troubleshooting Figure 3 18 Assembly A10 Exploded View 6691A 6692A ...
Page 60: ...Troubleshooting 57 Figure 3 19 Three Phase Line Choke Subchassis Wiring ...
Page 61: ...58 Troubleshooting Figure 3 20 24 Volt Fan Transformer ...
Page 77: ...74 Diagrams Figure 6 1 Test Point Waveforms for Table 6 3 sheet 2 of 2 ...
Page 79: ...Figure 6 3 A1 Front Panel Board Assembly Diagram ...
Page 81: ...Figure 6 5 A2 GPIB Board Component Location ...
Page 83: ...Figure 6 7 A3 FET Board Component and Test Point Location ...
Page 87: ...Figure 6 11 A4 AC Input Board Component and Test Point Location 12 9 10 11 ...
Page 89: ...Figure 6 13 A5 DC Rail Board Component and Test Point Location 13 14 ...
Page 91: ...Figure 6 15 A6 Bias Board Component and Test Point Location 15 16 17 18 19 19 20 22 ...
Page 94: ...Figure 6 17 Power Mesh Schematic Diagram All Models ...
Page 103: ......