Features and Functions
3
Keysight 34980A Mainframe User’s Guide
203
Factory Reset State
The following tables show the state of the instrument after a
*RST
or
SYSTem:CPON
command is executed.
Measurement Configuration
Factory Reset State
Function
Range
Resolution
Integration Time
Input Resistance
Channel Labels
Channel Delay
Reading Format
Sample Count
Trigger Count
Trigger Delay
Trigger Source
DC Volts
Autorange
5½ Digits
1 PLC
10 M
Ω
(fixed for all DCV ranges)
No Change
Automatic Delay
Reading Only (no units, channel, time)
1 Sample per Trigger
1 Trigger
Automatic Delay
Immediate
Scanning Operations
Factory Reset State
Scan List
Reading Memory
Min, Max, and Average
Sweep Count
Trigger Interval
Monitor in Progress
Empty
All Readings are Cleared
All Statistical Data is Cleared
1 Sweep
1 Second
Stopped
Mx+B Scaling
Factory Reset State
Scaling State
Gain Factor (“M”)
Offset Factor (“B”)
Scale Label
Off
1
0
Null String (“ “)