3
Features and Functions
158
Keysight 34980A Mainframe User’s Guide
the total number of readings returned will be the product of the three
parameters.
– The
CONFigure
and
MEASure?
commands automatically set the scan trigger
count to 1.
– The instrument sets the scan trigger count to 1 after a Factory Reset (
*RST
command). An Instrument Preset (
SYSTem:PRESet
command) or Card Reset
(
SYSTem:CPON
command) does
not
change the setting.
Front Panel Operation:
Scan (Configure) > SCAN TRIGGER >
COUNTED|INFINITE
Remote Interface Operation:
TRIGger:COUNt
To configure a continuous scan, send
TRIG:COUNT INFINITY
.
Sweep Count
The sweep count sets the number of sweeps per trigger event during a scan (a
sweep
is one pass through the scan list). The front-panel sample annunciator (“ * ”)
turns on during each measurement.
Sweep count
– The sweep count is valid only while scanning. If no channels have been
assigned to the scan list, the specified sweep count is ignored (no error is
generated).
t
Sweep 1
Trigger
Sweep Count
(1 to 500,000 sweeps)
Sweep 2
Sweep
n
Trigger
. . .