Features and Functions
3
Keysight 34980A Mainframe User’s Guide
199
Remote Interface Operation:
To read the count on either the specified internal
DMM relay or module channel relays, send the following commands.
DIAG:DMM:CYCLES? 2
DIAG:RELAY:CYCLES? (@1003,1013)
To reset the cycle count on the specified module channel relays, send the
following command (the instrument must be unsecured).
DIAG:RELAY:CYCLES:CLEAR (@1003,1911)
SCPI Language Version
The instrument complies with the rules and conventions of the present version of
SCPI (
Standard Commands for Programmable Instruments
). You can determine
the SCPI version with which the instrument is in compliance by sending a
command from the remote interface.
– You can query the SCPI version from the remote interface only.
– The SCPI version is returned in the form “YYYY.V”, where “YYYY” represents the
year of the version, and “V” represents a version number for that year (for
example, 1994.0).
Remote Interface Operation:
SYSTem:VERSion?