Functions and Measurements
Channel Scanner
68
Channel Scanner Setup
To customized the channel scan process, you can press
Meas
> Channel Scanner > Setup
to set the measurement interval and
rule in the submenu.
The
Meas Interval
is a value of time/distance during which the
every two scan process is triggered. When it is set to off, the
channel will be scanned with default settings.
The
Interval Type
is toggled between time and distance. When
it is set to
Time
, the minimum interval time is 100 ms. When
it is set to
Distance
, it allows you to measure the channel
with the specific distance on the road test. The minimum
interval distance is 1 meter.
The
Measure Rule
and
Search Rule
is toggled between
Speed
and
Accuracy
.
For each scan process, the analyzer will first search and find
out the available channels according to the selected scan
mode. Then, a precise measurement will perform to get the
scan result.
The accuracy mode in measurement rule is set as default to
get the precise measurement for the searching- out channels.
The speed mode will speed the measurement but reduce the
measurement accuracy. It is set to default in search rule to
search out the available channels approximately.
The
Top Cycle
number is the searching times for each
measurement. It allows you to measure the specific
searching- out channels continuously.
Summary of Contents for N9322C
Page 8: ...Contents 6 ...
Page 108: ...Functions and Measurements Files Operation 100 ...
Page 174: ...Key Reference Trace 166 ...
Page 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 210: ...Menu Maps Trace Det Menu 202 ...
Page 213: ......