Functions and Measurements
Analysis the Modulated Signals
92
FSK Demodulation Analysis
Please refer to the procedures below to analysis the FSK
demodulation.
1
Connect the signal generator to RF IN connector and turn
on the signal generator’s FSK modulation and RF output.
2
Turn on the ASK demodulation analysis function.
Press
MODE > Modulation Analysis
>
FSK
.
3
Press
Frequency > Carrier Freq > 100 > MHz
to set the carrier
frequency.
4
Press
Meas > Symbol Rate > 10 > ksps
to set the symbol rate.
5
Press
Filter Setup > Ref Filter > Nyquist
to set the filter type.
6
Set the attenuation state to Auto (default):
Press
Amplitude > Attenuation Auto
7
Setting the FSK deviation limit and turn on the limit:
Press
Shift > Limit > FreqDev UP > 85 > kHz > FreqDev Low > 80
> kHz > Limits on
The figure below shows the FSK symbol and the
measurement result.
Figure 3-32 FSK Symbol View