Functions and Measurements
Making Distortion Measurements
54
The signal produces harmonic products (spaced 200 MHz
from the original 200 MHz signal) in the analyzer input
mixer as shown in
.
4
Change the center frequency to the value of the first
harmonic:
Press
Peak Search > Next Peak
Press
Marker > Marker To > Mkr -> CF
5
Change the span to 50 MHz and re- center the signal:
Press
Span > 50 > MHz
Press
Peak Search
Press
Marker > Marker To > Mkr -> CF
6
Set the attenuation to 0 dB:
Press
Amplitude > Attenuation > 0 > dB
7
To determine whether the analyzer generates harmonic
distortion products, first display the trace data in trace 2 as
follows:
Press
Trace/Det > Select Trace > Trace 2
Press
Clear Write
8
Allow trace 2 to update (minimum two sweeps), then store
the data from trace 2 and place a delta marker on the
harmonic of trace 2:
Press
Trace/Det > View
Press
Peak Search
Press
Marker > Mode > Delta
The analyzer display shows the stored data in trace 2 and
the measured data in trace 1. The
Marker
Δ
indicator reads
the difference in amplitude between the reference and active
markers.
9
Increase the RF attenuation to 10 dB:
Press
Amplitude > Attenuation > 10 > dB
Summary of Contents for N9322C
Page 8: ...Contents 6 ...
Page 108: ...Functions and Measurements Files Operation 100 ...
Page 174: ...Key Reference Trace 166 ...
Page 206: ...Menu Maps Name Editor Menu 198 Name Editor Menu ABCDEFG HIJKLMN Return OPQRSTU VWXYZ Name ...
Page 208: ...Menu Maps SPAN Menu 200 SPAN Menu Span Span Full Span Zero Span Last Span ...
Page 210: ...Menu Maps Trace Det Menu 202 ...
Page 213: ......